• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. AMBIENT HIPVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. AMBIENT HIPVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number ASHA4730-01
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Foreign zip code: (b)(6).
 
Event or Problem Description
It was reported that during the surgery the electrode head detached from the electrode during ablation inside the patient, with the ask forceps everything was removed.No patient injury or other complications were reported.Another smith & nephew device was used to complete the surgery.
 
Additional Manufacturer Narrative
Single use device that was reprocessed and reused?: corrected.The returned device, intended for use in treatment, was returned as an mdr for evaluation.There was a relationship between the returned device and the reported event.Visual inspection under magnification shows extensive wear at distal end.The spacer with the screen/electrodes is completely detached.The detached spacer with attached screen/electrodes was returned for evaluation.The screen/electrodes show contamination but no abnormalities; there are no manufacturing abnormalities visually observed with the returned device.The device had no functionality and registered an error e-7 when plugged into a known good quantum2 controller.The suction line was tested and performed as intended.After bypassing the error e-7 the remaining part of the distal end of the device responded when ablate / coag was activated.The complaint was verified, however the root cause could not be determined with certainty since the device was comprehensively used.Excessive wear of the distal end results from vigorous use against bony surfaces; the wand is supplied sterile, and is for single use only.Do not clean, resterilize, or reuse the wand, as this may damage or compromise the performance resulting in product malfunction, failure, or patient injury.Cleaning, resterilization, or reuse of the wand may also expose the patient to the risk of transmitting infectious diseases or results in irregular wear leading to malfunction.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBIENT HIPVAC 50 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8332800
Report Number3006524618-2019-00092
Device Sequence Number11297090
Product Code GEI
UDI-Device Identifier00817470000436
UDI-Public00817470000436
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Catalogue NumberASHA4730-01
Device Lot Number2019198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Initial Date Received by Manufacturer 01/23/2019
Supplement Date Received by Manufacturer03/08/2019
Initial Report FDA Received Date02/12/2019
Supplement Report FDA Received Date03/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-