Model Number 7133
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/17/2018
Event Type  malfunction  
Event Description
It was reported that shaft break occurred.A 2.50mm x 15mm fg emerge balloon catheter was advanced for treatment.However, when the device was inserted, the shaft bent and came apart from the balloon when it was removed.The procedure was completed with another of the same device.No patient complications were reported.
 
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube, inflation lumen, and guidewire lumen.The hypotube is separated 80.9cm from the hub.Microscopic examination revealed no additional damages.The balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
It was reported that shaft break occurred.A 2.50mm x 15mm fg emerge balloon catheter was advanced for treatment.However, when the device was inserted, the shaft bent and came apart from the balloon when it was removed.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8332953
MDR Text Key135949246
Report Number2134265-2019-01164
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729806219
UDI-Public08714729806219
Combination Product (y/n)N
PMA/PMN Number
K113220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/02/2021
Device Model Number7133
Device Catalogue Number7133
Device Lot Number0022888660
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2019
Initial Date Manufacturer Received 01/23/2019
Initial Date FDA Received02/12/2019
Supplement Dates Manufacturer Received03/22/2019
Supplement Dates FDA Received04/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-