Catalog Number SMXP120RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was reviewed, and no anomalies were found related to this complaint.In addition, the mre review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) female patient underwent an unspecified procedure with an artoura breast tissue expander 375cc and experienced deflation on the left side.As a result, the patient underwent removal and replacement with an unspecified breast implant on (b)(6) 2019.
 
Additional Manufacturer Narrative
On 4/10/2019, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation summary: the analysis results show that the device appeared intact.Leak testing revealed there was a tear at the union between the dome union ( found in the anterior view) and shell of the implant.Microscopic examination of the edges of the rent gave no indications of instrument damage.No other anomalies observed.Microscopic examination of the edges of the tear was performed and it did not provide conclusive evidence of what could be the root cause.The condition observed on the returned sample might be related to an excessive force applied to the device, this cannot be conclusively determined, as rupture is a known complication associated with these devices and is referenced in the product insert data sheet.The manufacturing records were reviewed and the manufacturing criteria were met prior to the release of this lot.Deflation is a known complication associated with mentor mammary prostheses.Possible causes of deflation of saline-filled implants include, but are not limited to the following events: excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma to the breast.The reported complaint was confirmed.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Complaint information will be included in complaint trending that is reviewed and monitor by quality assurance on a regular basis to determine if further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8333378
Report Number1645337-2019-08500
Device Sequence Number1663713
Product Code LCJ
UDI-Device Identifier00081317024309
UDI-Public00081317024309
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2020
Device Catalogue NumberSMXP120RH
Device Lot Number7333723
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2019
Initial Date Received by Manufacturer 01/16/2019
Supplement Date Received by Manufacturer04/10/2019
05/22/2019
Initial Report FDA Received Date02/12/2019
Supplement Report FDA Received Date05/06/2019
05/25/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
-
-