| Catalog Number |
03.632.003 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that the patient underwent a matrix mis (minimally invasive surgery) spinal fusion procedure on (b)(6) 2019.It was found out that four (4) screwdrivers were all suffering from normal wear and tear.There was no surgical delay.Procedure outcome is unknown.There was no patient consequence.This report is for one (1) t25 stardrive shaft f/matrix cannulated/standard.This is report 1 of 4 for (b)(4).
|
| |
|