| Catalog Number |
RONYX35018W |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
02/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During the procedure an attempt was made to use a resolut onyx rx coronary drug eluting stent to treat a mildly tortuous and moderately calcified lesion located in the proximal rca with 90% stenosis.The device was inspected with no issues noted.Negative prep was not performed.The lesion was not predilated.It was reported that stent dislodgement occurred during delivery to the lesion.An attempt was made to remove the stent using a snare, however this was unsuccessful.A 4 x 20mm euphora balloon was used to deploy the stent.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the stent was not present on the balloon and did not return for analysis.The balloon folds remained intact.Crimp impressions were visible on the exposed balloon surface.The inner lumen patency was verified with a 0.015 inch mandrel.No damage was evident to the remainder of the delivery system.If information is provided in the future, a supplemental report will be issued.
|
| |
|