MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX OTW; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX OTW; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX35018W
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 02/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure an attempt was made to use a resolut onyx rx coronary drug eluting stent to treat a mildly tortuous and moderately calcified lesion located in the proximal rca with 90% stenosis.The device was inspected with no issues noted.Negative prep was not performed.The lesion was not predilated.It was reported that stent dislodgement occurred during delivery to the lesion.An attempt was made to remove the stent using a snare, however this was unsuccessful.A 4 x 20mm euphora balloon was used to deploy the stent.No patient injury was reported.
 
Additional Manufacturer Narrative
Device evaluation summary: the stent was not present on the balloon and did not return for analysis.The balloon folds remained intact.Crimp impressions were visible on the exposed balloon surface.The inner lumen patency was verified with a 0.015 inch mandrel.No damage was evident to the remainder of the delivery system.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX OTW
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8334577
Report Number9612164-2019-00496
Device Sequence Number1650358
Product Code NIQ
UDI-Device Identifier00643169602601
UDI-Public00643169602601
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2020
Device Catalogue NumberRONYX35018W
Device Lot Number0008950036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2019
Initial Date Received by Manufacturer 02/12/2019
Supplement Date Received by Manufacturer04/15/2019
Initial Report FDA Received Date02/13/2019
Supplement Report FDA Received Date04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-