• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX22522UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 02/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure an attempt was made to use a resolute onyx rx coronary drug eluting stent (2.25 x 22mm) to treat a mildly tortuous and moderately calcified lesion in the distal rca.There were no issues noted removing the device from the hoop.The lesion was pre-dilated.The device did pass through a previously-deployed stent.Resistance was encountered when advancing the device.Excessive force was used during delivery.There was a proximal lesion in the rca as well as a mid to distal lesion in the rca.The proximal lesion was pre-dilated and stented using 3.0 x 22mm resolute onyx, and post-dilated with an 3.5 nc euphora balloon.The distal lesion was then pre-dilated with a 2.0 euphora balloon and the 2.25 x 22mm stent was advanced through the proximal rca/stent to the mid-distal part of the rca.The stent could not be advanced far enough down the vessel for placement, and an attempt was made to pull back the stent to further pre-dilate the distal lesion before placing the stent.When the stent was pulled back, the stent became dislodged from the balloon when it got caught on the distal edge of the proximal 3.0 x 22 stent.A non medtronic balloon was used and was able to carry and advance the dislodged stent down the rca to the distal lesion.Multiple nc euphora balloons were then used in the stent to inflate and expand the dislodged stent and oppose it to the vessel wall.Another stent (2.25 x 12mm) was implanted over the dislodged stent at 20 atmospheres.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8334610
Report Number9612164-2019-00498
Device Sequence Number1691984
Product Code NIQ
UDI-Device Identifier00643169556737
UDI-Public00643169556737
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2020
Device Catalogue NumberRONYX22522UX
Device Lot Number0009379540
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2019
Initial Report FDA Received Date02/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-