| Model Number |
1032800 |
| Medical Device Problem Codes |
Failure to Recalibrate (1517); Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The device has not yet been returned to the manufacturer.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported a ventilator was alarming for a ventilator inoperative alarm condition.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's software was reloaded to address the issue.During the evaluation of the device at the manufacturer's service center, the device failed a test step during testing.The device's active exhalation control module was replaced to address the issue.
|
| |
|