Catalog Number T5C4600 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that "all lines" were disconnected from the kaguya set.This occurred after a demonstration on the kaguya machine was performed by the nurse.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection was performed and it was noted that all the lines had separated from the welded cassette subassembly.Solvent residue was present in the tubing and the tubing showed evidence that it had been inserted fully into the port during the manufacturing process.An underwater pressure test was also performed with no issued noted.The reported condition was verified.The cause of the reported condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|