MAUDE Adverse Event Report: BAXTER HEALTHCARE - SINGAPORE HOME PD SYSTEM KAGUYA SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - SINGAPORE HOME PD SYSTEM KAGUYA SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number T5C4600
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 11/16/2018
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that "all lines" were disconnected from the kaguya set.This occurred after a demonstration on the kaguya machine was performed by the nurse.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection was performed and it was noted that all the lines had separated from the welded cassette subassembly.Solvent residue was present in the tubing and the tubing showed evidence that it had been inserted fully into the port during the manufacturing process.An underwater pressure test was also performed with no issued noted.The reported condition was verified.The cause of the reported condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOME PD SYSTEM KAGUYA SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
MDR Report Key8337093
MDR Text Key136114291
Report Number1416980-2019-00795
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberT5C4600
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Initial Date Manufacturer Received 01/16/2019
Initial Date FDA Received02/13/2019
Supplement Dates Manufacturer Received02/18/2019
Supplement Dates FDA Received03/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-