CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 190618 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
A user facility reported a fresenius 2008k2 machine that powered off and gave off a burning smell.The issue occurred prior to any treatment, and there was no patient involvement.Upon follow up, it was confirmed that the machine had melted fuses that were found during repair.A fresenius (b)(4) technician replaced the burned fuses and the power switch to resolve the issue and return the machine to service.Additional information was requested, but was not provided.If additional information is provided, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius mexico service technician, who reportedly repaired the machine by replacing the burned fuses and repairing the damaged power switch.The machine passed functional tests and returned to service.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
|
|
Search Alerts/Recalls
|
|
|