Catalog Number 190618
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 01/16/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility reported a fresenius 2008k2 machine that powered off and gave off a burning smell.The issue occurred prior to any treatment, and there was no patient involvement.Upon follow up, it was confirmed that the machine had melted fuses that were found during repair.A fresenius (b)(4) technician replaced the burned fuses and the power switch to resolve the issue and return the machine to service.Additional information was requested, but was not provided.If additional information is provided, a supplemental report will be submitted.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius mexico service technician, who reportedly repaired the machine by replacing the burned fuses and repairing the damaged power switch.The machine passed functional tests and returned to service.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8339028
MDR Text Key136126920
Report Number2937457-2019-00498
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190618
Was Device Available for Evaluation? Yes
Device Age MO
Initial Date Manufacturer Received 02/12/2019
Initial Date FDA Received02/14/2019
Supplement Dates Manufacturer Received03/01/2019
Supplement Dates FDA Received03/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-