C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER; X- FORCE Back to Search Results
Model Number 998006
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Codes No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the lower end of the balloon ruptured at less than 20 atmospheres when the doctor injected the contrast agent into the balloon.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿method of use: 1.Do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended rbp.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.2.Do not use air or any gaseous substances as a balloon inflation media, always use sterile liquid media.3.If resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation." the device was not returned.
 
Event or Problem Description
It was reported that the lower end of the balloon ruptured at less than 20 atmospheres when the doctor injected the contrast agent into the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
X- FORCE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8339752
Report Number1018233-2019-00782
Device Sequence Number19396454
Product Code EZN
UDI-Device Identifier10801741052801
UDI-Public(01)10801741052801
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model Number998006
Device Catalogue Number998006
Device Lot NumberBMBWFM44
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/22/2019
Supplement Date Received by Manufacturer03/28/2019
Initial Report FDA Received Date02/14/2019
Supplement Report FDA Received Date04/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-