| Catalog Number |
L20416 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/ explanted.Examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Territory 226 submitted a complaint with an unknown reason.Examination of the returned instrument found deep circular scratching on the taper.A functional check with a mating broach handle found the circular scratching on the taper prevented from seating correctly into the broach handle.The root cause is attributed to misuse.Deep circular scratching is due to the taper not being seated all the way into the broach handle before locking down on the lever locking mechanism of the broach handle.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.
|
| |
|
Event or Problem Description
|
|
It was reported that the instrument office sets had an unspecified malfunction.It did not affect surgery or patient.In-house engineering evaluation determined the following: a functional check with a mating broach handle found the circular scratching on the taper prevented from seating correctly into the broach handle.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|