| Model Number |
MT22719 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2019 that on (b)(6) 2019 the receiver ceased to function.No additional event or patient information is available.No product or data was provided for evaluation.The confirmation of the complaint is undetermined.The probable cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver ceased to function.The product was evaluated.An external visual inspection was performed and passed.Charge and boot testing was performed and passed.The log was downloaded and reviewed finding rtt loss within the investigation window.Receiver functional test was performed and passed.The receiver was opened and an internal visual inspection was performed and failed due to damaged battery/ cracked solder on battery ic pins.The allegation was confirmed.The probable cause was determined to be receiver damaged battery.No injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|