Model Number US35130515
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 12/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 02/2020), (manufacturing review: 04/2018).The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of a clinical trial, that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter, vessel dissection was identified after one inflation in the left superficial femoral artery.There was no reported treatment of the dissection.The current stats of the patient was not provided.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive, as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported through the results of a clinical trial, that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter, vessel dissection was identified after one inflation in the left superficial femoral artery.There was no reported treatment of the dissection.The current stats of the patient was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE FOCUSED FORCE PTA BALLOON
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8341429
Report Number2020394-2019-00129
Device Sequence Number11251465
Product Code PNO
UDI-Device Identifier00801741116797
UDI-Public(01)00801741116797
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS35130515
Device Catalogue NumberUS35130515
Device Lot NumberGFCP0624
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2019
Supplement Date Received by Manufacturer04/05/2019
Initial Report FDA Received Date02/14/2019
Supplement Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age71 YR
Patient Weight65
-
-