| Model Number |
MMT-551LNAB |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The insulin pump was received with intermittent button response due to flattened dome switch on esc and act button.The connector was inspected and locked on lcd board.No button error alarm during testing.The insulin pump passed displacement test and self test.(b)(4).The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump buttons were not responding.The customer¿s blood glucose level was 145 mg/dl at the time of the incident.Customer was advised to observe time clock for one minute to verify if time advances and they stated that the time was advances.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The date provided with the initial report was incorrect.The correct date is (b)(4) 2019.
|
| |
|