MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAL
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The insulin pump was received with intermittent button response due to flattened dome switch on up arrow and down arrow button.The connector was inspected and locked on lcd board.No button error alarm during testing.Pump passed displacement test and self test.(b)(4).The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer reported via phone call the insulin pump had a keypad anomaly.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump up and down buttons were less responsive.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
The date provided with the initial report was incorrect.The correct date is (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8341904
Report Number3004209178-2019-55881
Device Sequence Number13941739
Product Code OYC
UDI-Device Identifier00643169693791
UDI-Public(01)00643169693791
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAL
Device Catalogue NumberMMT-723NAL
Device Lot NumberB6723NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer02/25/2019
Initial Report FDA Received Date02/14/2019
Supplement Report FDA Received Date02/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight195
-
-