| Model Number |
MMT-723NAL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The insulin pump was received with intermittent button response due to flattened dome switch on up arrow and down arrow button.The connector was inspected and locked on lcd board.No button error alarm during testing.Pump passed displacement test and self test.(b)(4).The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had a keypad anomaly.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump up and down buttons were less responsive.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The date provided with the initial report was incorrect.The correct date is (b)(6) 2019.
|
| |
|