The reported event was confirmed as manufacturing related.Visual evaluation noted one unopened hydrophilic person catheter.The seal was intact all the way around.The water packet had some type of damage on one side of the packaging, it was unclear if that allowed any of the water to escape the packaging, but there was no evidence of water in the surrounding packaging or catheter.The water packet was removed from the catheter packaging, and inspected for holes or leaks.No holes or leaks were found.There was still some water in the packet, but not enough to burst the packet.The water packet was cut and the contents were poured into a graduated cylinder.The water volume measured to be 1.3ml which was found to be out of specification (4ml +/- 0.3ml).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿for urological use only.Intended for use by patients for bladder management including urine drainage, collection and measurement.The devices are passed to the urinary bladder via the urethra.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/precautions and adverse reactions.This is a single use device.Do not resterilize.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.1.Wash your hands thoroughly with soap and water.2.Prior to opening the sealed catheter pouch, apply pressure to the foil packet to release the water.Ensure all water is released from the foil packet.3.Wet the catheter by holding the package with the printed side up and tip the package end-to-end three to six times to wet the catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.4.Peel open the pack at the funnel/handle end just enough to expose 5 cm (2¿) of the catheter, or for products with an insertion sleeve, expose the insertion sleeve.Don¿t remove the catheter yet.Use the adhesive tab at the funnel end of the pack to stick the pack to a nearby vertical surface while preparing to catheterize.5.Wash the area around the meatus before catheterizing.6.Wash your hands again.7.Using the catheter with insertion sleeve (gripper): a.Hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.B.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.C.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Advance the catheter until urine starts to flow.Without insertion sleeve (gripper): a.Hold the handle/funnel end and remove the catheter from the packaging.B.Gently pass the tip of the catheter into your urethra and advance the catheter until urine starts to flow.8.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.9.Finish by disposing of the catheter and its packaging.Wash your hands with soap and water.".
|