| Catalog Number |
03L79-31 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Event or Problem Description
|
|
The customer reported a false elevated calcium result when processing on the architect c8000.The initial result was 140 mg/l (14mg/dl).Retest yielded a result in normal range (no specific value provided).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed for the customer issue and included a review of the complaint text, a search for similar complaints, tracking and trending, a review of historical performance and product labeling.A search by product lot number for other tickets similar to this issue found no similar complaints.The trend review by the product list number found no adverse or non-statistical trends related to this issue.World wide data was used to review historical performance and no unusual reagent lot performance was identified for lot 32860un18.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation a product deficiency was not identified.
|
| |
|
Search Alerts/Recalls
|