• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT CEM TIBIAL TRAY SZ2.5; MBT TIBIAL TRAY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT CEM TIBIAL TRAY SZ2.5; MBT TIBIAL TRAY : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 129431125
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
 
Event or Problem Description
The patient was revised to address loosening of the tibial component at the cement to implant interface.Unknown cement was used.Doi: (b)(6) 2008; dor: (b)(6) 2018; left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT CEM TIBIAL TRAY SZ2.5
Common Device Name
MBT TIBIAL TRAY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8342869
Report Number1818910-2019-84657
Device Sequence Number13753952
Product Code NJL
UDI-Device Identifier10603295025597
UDI-Public10603295025597
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number129431125
Device Lot Number2462025
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/28/2019
Supplement Date Received by Manufacturer03/20/2019
Initial Report FDA Received Date02/15/2019
Supplement Report FDA Received Date03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-