| Catalog Number |
10400 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer in (b)(6) reported a non conforming quality control strain in association with the api® nh (neisseria gonorrhoeae) test strip (lot 100678088).The customer stated that their qc organism (haemophilus influenzae atcc® 10211) was conforming on a previous lot of api test strips, but not on lot 1006780880.The customer then tested a new atcc® 10211 qc organism and lot 100678088 was not conforming.The customer switched to a different api nh lot and the issue was resolved.The atcc® 10211 qc strain is used to perform comprehensive quality control to demonstrate positive and negative reactivity with apnh.A biomérieux internal investigation will be initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
A biomérieux internal investigation was conducted following a report of qc failure for haemophilus influenzae atcc® 10211¿ in association with the api nh test.1.Retained samples analysis.The conformity of the implicated lot 1006780880 and an internal reference lot 10066598040 used for the investigation were confirmed with the following atcc quality control strains tested for the release of this reference: atcc 31426 neisseria gonorrhoeae.Atcc 10211 haemophilus influenzae.Atcc 49917 haemophilus paraphrophilus.The results obtained for these strains with both lots are in agreement with the quality control specifications.2.Complaints trend analysis.No other complaint was registered on the lot 1006780880.No investigation was opened in the past for qc failed on this reference.3.Conclusion.The results obtained with the atcc quality control strains tested during the investigation conform with specifications.The issue observed by the customer was not reproduced on the retained sample tested.The investigation concluded that the api nh test kit is performing as intended.
|
| |
|
Search Alerts/Recalls
|