| Catalog Number |
1011493-15 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Date of event - estimated.The device was not returned for analysis.A review of the lot history record and complaint history could not be conducted because the lot number was not provided.The investigation was unable to determine a conclusive cause for the difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported the procedure was to treat a lesion in the renal artery.The 5.0 mm x 15 mm x 135 cm herculink elite stent delivery system (sds) was being advanced to the target lesion and in an attempt to reposition the device; the sds was retracted and resistance was felt with the guide catheter.The stent was not implanted.The sds and guide catheter were removed together as a single unit.A new herculink elite stent was implanted to complete the procedure.There was no reported adverse patient effect or a clinically significant delay in the procedure.No additional information was provided.
|
| |
|