| Model Number |
MMT-723RNAL |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device received motor error alarm during rewind due to corroded motor home switch.Unable to perform displacement test, basic occlusion, occlusion, prime/compromised force sensor system and excessive no delivery test due to motor error alarm during the rewind test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump received motor error.The customer¿s blood glucose level was unknown.The customer was advised the pump will need to be replaced.The customer was advised to discontinue use of the pump and was advised to refer to back up plan, per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The date provided in the initial report was incorrect.The correct date is (b)(4) 2019.
|
| |
|