| Model Number |
PHYSICIAN CHOICE ENCORE TOWER |
| Medical Device Problem Code |
Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The suspected device was received and a failure investigation was conducted.The reported issue was confirmed and a damaged (burnt) contact was found on the isolation transformer's switch.This issue will be internally investigated within vyaire.
|
| |
|
Event or Problem Description
|
|
The customer initially reported the static and dynamic compliance vmax device was unable to power on.The customer in conjunction with the field service representative determined the most likely caused of event vmax cart.Customer reported there was no patient involvement with the event.
|
| |
|