Model Number 172524-LVISJ
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The pusher and carotid stent delivery system was returned for analysis.Upon initial investigation, the distal tip of the pusher appeared to have been damaged, and the platinum marker detached stainless steel coils are not present at the distal tip of the pusher.The distal marker band to the distal tip of the pusher was measured to see how much was detached; approximately 17mm remained, thus meaning that approximately 20mm of the pusher was detached.Based on the available information and investigation, the reported complaint is confirmed; the failure mode of the pusher was consistent with a kink overload resulting in the fracture.The broken end was severely bent backwards upon itself and twisted, which is consistent with it being kinked.The root cause of this complaint is due to the wire being twisted into a kink and then pulled away causing a shear/tensile overload.
 
Event or Problem Description
It was reported that during treatment of a left-side, posterior communicating aneurysm, the tip of the delivery wire of the lvis jr stent detached in the internal carotid artery at the neck of the blister aneurysm.This was noticed after the remainder of the lvis stent was re-sheathed and removed from the body.Another lvis jr.Stent was then placed to secure the detached segment of the stent to the wall of the vessel.There was no reported injury and the patient was noted to be in fair condition post-procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LVIS JR 3.5X23
Common Device Name
STENT
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key8346185
Report Number2032493-2019-00028
Device Sequence Number19290981
Product Code NJE
UDI-Device Identifier00842429100844
UDI-Public(01)00842429100844(11)180509(17)210430(10)18050953EM
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K130005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number172524-LVISJ
Device Catalogue Number172524-LVISJ
Device Lot Number18050953EM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2019
Initial Report FDA Received Date02/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
-
-