• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY TIMESH 1 X PLATE, STRAIGHT; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY TIMESH 1 X PLATE, STRAIGHT; PLATE, BONE Back to Search Results
Model Number 015-044
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One timesh plate was returned.One of the end eyelets was broken and deformed leaving a gap through the aperture.It is unknown how this damage occurred.The instructions for use cautions, ¿excess force, over-bending, notching, or scratching during mesh forming process could result in mesh fracture.¿ no other unusual characteristics were noted.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the skull repair surgery, the plate broke, and the screw broke in the middle.The products were replaced and there was no injury to the patient.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIMESH 1 X PLATE, STRAIGHT
Common Device Name
PLATE, BONE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key8346250
Report Number2021898-2019-00056
Device Sequence Number1058068
Product Code JEY
UDI-Device Identifier00673978002682
UDI-Public00673978002682
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973145
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number015-044
Device Catalogue Number015-044
Device Lot NumberT173885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2019
Supplement Date Received by Manufacturer02/22/2019
Initial Report FDA Received Date02/15/2019
Supplement Report FDA Received Date03/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
Patient Weight52
-
-