| Model Number |
015-044 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One timesh plate was returned.One of the end eyelets was broken and deformed leaving a gap through the aperture.It is unknown how this damage occurred.The instructions for use cautions, ¿excess force, over-bending, notching, or scratching during mesh forming process could result in mesh fracture.¿ no other unusual characteristics were noted.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the skull repair surgery, the plate broke, and the screw broke in the middle.The products were replaced and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|