| Catalog Number |
RONYX22538UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
02/15/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use one resolute onyx rx coronary drug eluting stent during a procedure.The device was inspected with no issues.Negative prep was performed with no issues.It was reported that stent dislodgement occurred during delivery to/at the lesion.The dislodged stent was removed using a snare.
|
| |
|
Additional Manufacturer Narrative
|
|
There was not a lot of tortuosity or calcification at the lesion.It was indicated that it was not a challenging lesion.No difficulties were noted when removing the protective sheath.The stent was inspected post removal of the protective sheath with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent or cell of a stent.No significant resistance was noted when advancing the device.Excessive force was not used during delivery.If information is provided in the future, a supplemental report will be issued.
|
| |
|