• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX22538UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 02/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one resolute onyx rx coronary drug eluting stent during a procedure.The device was inspected with no issues.Negative prep was performed with no issues.It was reported that stent dislodgement occurred during delivery to/at the lesion.The dislodged stent was removed using a snare.
 
Additional Manufacturer Narrative
There was not a lot of tortuosity or calcification at the lesion.It was indicated that it was not a challenging lesion.No difficulties were noted when removing the protective sheath.The stent was inspected post removal of the protective sheath with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent or cell of a stent.No significant resistance was noted when advancing the device.Excessive force was not used during delivery.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8347034
Report Number9612164-2019-00534
Device Sequence Number1867136
Product Code NIQ
UDI-Device Identifier00643169556775
UDI-Public00643169556775
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2020
Device Catalogue NumberRONYX22538UX
Device Lot Number0009163133
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/15/2019
Supplement Date Received by Manufacturer02/25/2019
Initial Report FDA Received Date02/18/2019
Supplement Report FDA Received Date03/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-