DEPUY ORTHOPAEDICS, INC. 1818910 SROM*STM STD,36NK,20X15X165; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 523194
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Scar Tissue (2060); Tissue Damage (2104); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191)
Date of Event 07/22/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Initial reporter occupation: non-healthcare professional "attorney".(b)(4).
 
Event or Problem Description
Litigation papers allege physical pain, bodily impairment, debilitating lack of mobility, and high levels of toxic metal in her blood stream.Update sales rep reported revision surgery.Patient was revised to address pain.Update in addition to what were previously alleged, ppf alleges metallosis and metal wear.After review of medical records, the patient was revised to address left failed metal on metal hip arthroplasty.It was also confirmed that the patient had elevated metal ions.Revision notes indicated that the trunnion was noted to have metallosis.Doi: (b)(4) 2008 - dor: (b)(4) 2014, (left hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database ve:search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
MDR Report Key8347296
Report Number1818910-2019-84834
Device Sequence Number16623154
Product Code JDI
UDI-Device Identifier10603295170884
UDI-Public10603295170884
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K851422
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2012
Device Catalogue Number523194
Device Lot Number2339808
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/04/2019
Supplement Date Received by Manufacturer10/09/2019
Initial Report FDA Received Date02/18/2019
Supplement Report FDA Received Date10/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient Weight84
-
-