STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
502-01-52E |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Codes |
Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
01/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding shell malposition and rom involving a trident shell was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: could not be performed as lot code information was not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including device details, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to dislocation, subluxation and cup impingement.Intra-operatively, it was observed that the shell was well-fixed but malpositioned, the lip of the liner was found to be grooved, worn, and cracked, and mild wear debris was found in the head-trunnion interface.A trident hemispherical cluster hole shell, 10° e trident liner, and 36 +5 v40 lfit head were revised to a tritanium revision shell with 4 screws, a 40 liner, and a 40 metal head.The accolade i stem was well-fixed with the trunnion in good condition and was not revised.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to dislocation, subluxation and cup impingement.Intra-operatively, it was observed that the shell was well-fixed but malpositioned, the lip of the liner was found to be grooved, worn, and cracked, and mild wear debris was found in the head-trunnion interface.A trident hemispherical cluster hole shell, 10° e trident liner, and 36 +5 v40 lfit head were revised to a tritanium revision shell with 4 screws, a 40 liner, and a 40 metal head.The accolade i stem was well-fixed with the trunnion in good condition and was not revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding shell malposition and rom involving a trident shell was reported.The event was not confirmed.Method & results: -device evaluation and results: visual inspection visual inspection was performed as part of the material analysis report (mar), dated 29 mar 2019.This inspection indicated.[.] biological fixation was observed on the shell, [.].Damage was observed on the distal surface of the shell, consistent with contact against a hard object -medical records received and evaluation: no medical records were received for review with a clinical consultant¿¿ -device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.¿ additional information including , operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|