• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 502-01-52E
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 01/22/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding shell malposition and rom involving a trident shell was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: could not be performed as lot code information was not provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including device details, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
It was reported that the patient's right hip was revised due to dislocation, subluxation and cup impingement.Intra-operatively, it was observed that the shell was well-fixed but malpositioned, the lip of the liner was found to be grooved, worn, and cracked, and mild wear debris was found in the head-trunnion interface.A trident hemispherical cluster hole shell, 10° e trident liner, and 36 +5 v40 lfit head were revised to a tritanium revision shell with 4 screws, a 40 liner, and a 40 metal head.The accolade i stem was well-fixed with the trunnion in good condition and was not revised.
 
Event or Problem Description
It was reported that the patient's right hip was revised due to dislocation, subluxation and cup impingement.Intra-operatively, it was observed that the shell was well-fixed but malpositioned, the lip of the liner was found to be grooved, worn, and cracked, and mild wear debris was found in the head-trunnion interface.A trident hemispherical cluster hole shell, 10° e trident liner, and 36 +5 v40 lfit head were revised to a tritanium revision shell with 4 screws, a 40 liner, and a 40 metal head.The accolade i stem was well-fixed with the trunnion in good condition and was not revised.
 
Additional Manufacturer Narrative
Reported event: an event regarding shell malposition and rom involving a trident shell was reported.The event was not confirmed.Method & results: -device evaluation and results: visual inspection visual inspection was performed as part of the material analysis report (mar), dated 29 mar 2019.This inspection indicated.[.] biological fixation was observed on the shell, [.].Damage was observed on the distal surface of the shell, consistent with contact against a hard object -medical records received and evaluation: no medical records were received for review with a clinical consultant¿¿ -device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.¿ additional information including , operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL CLUSTER 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8347692
Report Number0002249697-2019-01075
Device Sequence Number16623150
Product Code LPH
UDI-Device Identifier07613153073086
UDI-Public07613153073086
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2009
Device Catalogue Number502-01-52E
Device Lot Number11418701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Initial Date Received by Manufacturer 01/22/2019
Supplement Date Received by Manufacturer04/08/2019
Initial Report FDA Received Date02/18/2019
Supplement Report FDA Received Date05/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight77
-
-