|
COVIDIEN MFG DC BOULDER LIGASURE IMPACT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Model Number |
LF4318 |
| Medical Device Problem Code |
Energy Output Problem (1431)
|
| Health Effect - Clinical Codes |
Blood Loss (2597); No Code Available (3191)
|
| Date of Event |
01/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during cystoprostatectomy, the device was able to activate normally and produced end tone, but the physician could not stop bleeding.Hemostasis was performed as the following steps: the use of the device to stop bleeding and when hemostasis failed, they used a different device to stop bleeding, but failed to stopped bleeding either, unit they used the method of soft coagulation.Also, the activation setting was changed to bar 3 from bar 2, however, the method did not have effect.Blood transfusion was done to patient at the end of the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily, and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|