| Catalog Number |
5532-G-709 |
| Medical Device Problem Codes |
Crack (1135); Fracture (1260)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
01/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right knee was revised due to instability.Intraoperatively, the post on the ps insert was noted to be broken.A 7x9 ps insert was revised to a 7x11 ps insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding instability due to ps insert post fracture was reported.The event was confirmed based on the images provided.Method & results: -device evaluation and results: the reported device was not returned however photographs were provided for review.An intraoperative image shows the insert in-situ, it can be clearly identified that the ps post has fractured from the insert.Another image shows the fractured post after explantation.The other images show damage to the underside of the insert component with what appears to be explantation damage.-medical records received and evaluation: no medical records were received for review with a clinical consultant -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: while the fracture of the ps insert post could be confirmed based on the photographs provided, the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right knee was revised due to instability.Intraoperatively, the post on the ps insert was noted to be broken.A 7x9 ps insert was revised to a 7x11 ps insert.
|
| |
|