Model Number G31519
Device Problem Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2019
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) # k083330.(b)(4).The device related to this occurrence underwent a laboratory evaluation.A proximal kink was observed during the lab evaluation.Prior to distribution, all echo-25 devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at (b)(4).A review of the manufacturing records for echo-25 of lot number c1505525 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1505525.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle kinked/bent was concluded from the available information.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to excessive force.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
It is impossible to put the stylet back into the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE
Type of Device
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
heather ryan
o halloran road
national technology park
limerick 
061334440
MDR Report Key8349228
MDR Text Key136613152
Report Number3001845648-2019-00065
Device Sequence Number1
Product Code FCG
UDI-Device Identifier00827002315198
UDI-Public(01)00827002315198(17)210601(10)C1505525
Combination Product (y/n)N
PMA/PMN Number
K083330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2021
Device Model NumberG31519
Device Catalogue NumberECHO-25
Device Lot NumberC1505525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/19/2019
Event Location Hospital
Initial Date Manufacturer Received 01/22/2019
Initial Date FDA Received02/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-