| Model Number |
6944-65 |
| Medical Device Problem Codes |
Intermittent Capture (1080); Failure to Capture (1081)
|
| Health Effect - Clinical Code |
Vomiting (2144)
|
| Date of Event |
01/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was possible intermittent biventricular capture and the intermittent biventricular pacing appeared not to capture.It was further reported the patient was vomiting.The leads remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated the patient was admitted with hyperkalemia and in diabetic ketoacidosis.A device check was later performed and noted normal function.The loss of capture was suspected to be related to the patient's abnormal lab results.
|
| |
|