MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  IN-CONDL CHISEL 8X10X7 3/4 IN; KNEE INSTRUMENT : CUTTING INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  IN-CONDL CHISEL 8X10X7 3/4 IN; KNEE INSTRUMENT : CUTTING INSTRUMENTS Back to Search Results
Catalog Number 242101000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.Device is an instrument and is not implanted/explanted.
 
Event or Problem Description
It was reported that the offset osteotome broke off during removal of tibial insert.There was no surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary :no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No device was received for examination, therefore the reported event could not be confirmed.A worldwide complaint database search shows similar reported events of device breakage.Additional reported event (com 219215) was investigated and found the tips of the chisels were loaded beyond the material limit resulting in mixed mode ductile and brittle overload of the material and fracture.The investigation could not draw any conclusions about the current reported event without the device to examine.Based on the inability to determine a root cause, a need for corrective action is not indicated.Monitor complaints through sep-419.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN-CONDL CHISEL 8X10X7 3/4 IN
Common Device Name
KNEE INSTRUMENT : CUTTING INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8349938
Report Number1818910-2019-85008
Device Sequence Number14024637
Product Code FZO
UDI-Device Identifier10603295122715
UDI-Public10603295122715
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242101000
Device Lot NumberSO2011364
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2019
Supplement Date Received by Manufacturer03/28/2019
Initial Report FDA Received Date02/19/2019
Supplement Report FDA Received Date03/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-