| Model Number |
U35130530 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Blood Loss (2597)
|
| Date of Event |
01/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number for the device was not provided; therefore, a review of the device history records could not be performed.The return of the device is pending.The investigation of the reported event is currently underway.
|
| |
|
Event or Problem Description
|
|
It was reported that at the completion of an angioplasty procedure in the superficial femoral artery, resistance was allegedly felt when removing the guidewire.The pta balloon and guidewire were removed as a single unit, which resulted in a tear at the access site the left common femoral artery.It was further reported that there was bleeding at the access site, for about an hour.The bleeding was stopped with astriction and no further treatment was needed.There were no further complications to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that at the completion of an angioplasty procedure in the superficial femoral artery, resistance was allegedly felt when removing the guidewire.The pta balloon and guidewire were removed as a single unit, which resulted in a tear at the access site the left common femoral artery.It was further reported that there was bleeding at the access site, for about an hour.The bleeding was stopped with astriction and no further treatment was needed.There were no further complications to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number for the device was not provided; therefore, a review of the device history records could not be performed.The device has been returned for evaluation.The investigation of the reported event is currently underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a complete manufacturing review could not be conducted, as the lot number was not reported for this device.Investigation summary: the device was returned for evaluation.No visual anomalies were noted with the returned device.An in-house 0.035" guidewire was able to pass without issue through the returned catheter.Therefore, the investigation is unconfirmed for the alleged guidewire issues.The definitive root cause for the reported rupture could not be determined based upon the available information.Label review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that at the completion of an angioplasty procedure in the superficial femoral artery, resistance was allegedly felt when removing the guidewire.The pta balloon and guidewire were removed as a single unit, which resulted in a tear at the access site the left common femoral artery.It was further reported that there was bleeding at the access site, for about an hour.The bleeding was stopped with astriction and no further treatment was needed.There were no further complications to the patient.
|
| |
|