• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN V-LOC; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN V-LOC; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number VLOCN0605
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic hernia, while closing the defect, the thread broke when pulling on the suture after the second stitch.New suture was used to complete the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of four devices, two sealed with the appropriate packaging and two opened by the account.The sealed products were representative samples.The visual inspection of the opened sample noted two needles, one suture and a partial suture.The partial suture was returned broken and missing the loop end of the suture.It was observed, under magnification, that the sutures appeared to have split under tension.Visual inspection and functional testing of the sealed products, representative samples, confirmed there were no abnormalities that would have caused or contributed to the reported condition.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.A definitive root cause could not be determined with regard to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-LOC
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mc dermott rd
north haven CT 06473
MDR Report Key8352111
Report Number1219930-2019-00963
Device Sequence Number207101
Product Code GAT
UDI-Device Identifier20884521171074
UDI-Public20884521171074
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K103052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberVLOCN0605
Device Catalogue NumberVLOCN0605
Device Lot NumberA8J0160VX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Initial Date Received by Manufacturer 01/24/2019
Supplement Date Received by Manufacturer04/12/2019
Initial Report FDA Received Date02/19/2019
Supplement Report FDA Received Date05/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-