| Catalog Number |
NSLG2S45 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #r9412r.Investigation summary: the device was received the upper jaw damaged at the proximal side.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The condition of the jaws prevented the functionality of the jaw.The device was unable to cycle open and close.Therefore we were unable to investigate further the reported alert screen.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during unknown procedure they are receiving the error code multiple times and they are switching them out.Ensealg2.Case completed with another device of the same product code.There were no patient consequences reported.
|
| |
|