| Model Number |
MMT-551NAH |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unit received with broken reservoir tube lip and missing reservoir tube o-ring.Unable to perform displacement test due to broken reservoir tube lip.A test reservoir was installed and did not lock in place due to broken reservoir tube lip.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that insulin pump had cosmetic damage.The blood glucose at the time of the incident was unknown.The customer was assisted with troubleshooting.The device will be returned for analysis.
|
| |
|