Model Number 110880
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code Extubate (2402)
Date of Event 01/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's age is 60-80 years old.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, a moisture was found in the pilot balloon, inflation line and cuff of the device.It was stated a sound was heard which was due to leak.The customer reported patient was reintubated.
 
Additional Manufacturer Narrative
Evaluation summary: one sample was returned for evaluation.Visual examination shows that the returned unit is cropped at the 23cm depth marking and there is some white sticky tape attached to the tubing at the 18th & 19th depth markings which is not from our manufacturing facility.Further examination shows that there is no fluid contained in the cuff body, the inflation line or the pilot balloon.Further examination also shows a small amount of air contained in the cuff body.The sample was inflated /deflated three times and no issue noted.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key8352972
Report Number8020889-2019-00019
Device Sequence Number2010235
Product Code BTR
UDI-Device Identifier40884521068231
UDI-Public40884521068231
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K082520
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Model Number110880
Device Catalogue Number110880
Device Lot Number201804214X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Initial Date Received by Manufacturer 01/22/2019
Supplement Date Received by Manufacturer05/07/2019
Initial Report FDA Received Date02/20/2019
Supplement Report FDA Received Date05/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-