• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL, CHECKPOINT, STERILE; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL, CHECKPOINT, STERILE; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 111651
Medical Device Problem Codes No Apparent Adverse Event (3189); Non Reproducible Results (4029)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
During a mako tka case that my colleague (b)(6) and i were running together, after all cuts were done and implants were cemented, the scrub nurse noticed that the tibial checkpoint was missing from the patient¿s tibia.Since the checkpoint could not be found anywhere, the surgeon requested an x-ray of the patient before closing the wound.At this stage, the checkpoint was found underneath the implanted tibial component, which at that point was already cemented on bone.
 
Event or Problem Description
During a mako tka case that my colleague shane madigan and i were running together, after all cuts were done and implants were cemented, the scrub nurse noticed that the tibial checkpoint was missing from the patient¿s tibia.Since the checkpoint could not be found anywhere, the surgeon requested an x-ray of the patient before closing the wound.At this stage, the checkpoint was found underneath the implanted tibial component (see attached x-ray picture), which at that point was already cemented on bone.
 
Additional Manufacturer Narrative
Reported event: it was reported that the checkpoint was left in the patient.Product evaluation and results: the product was unavailable for inspection as the product was not returned.Product history review: product history review was not performed as the lot number was not provided.Complaint history review: complaint history review was not performed as the lot number was not provided.Conclusions: the failure could not be determined as the product was not available for inspection.No additional investigation or specific actions are required.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL, CHECKPOINT, STERILE
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8353291
Report Number3005985723-2019-00159
Device Sequence Number16664525
Product Code OLO
UDI-Device Identifier00848486001353
UDI-Public00848486001353
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K112507
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number111651
Device Lot NumberN/A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/17/2019
Supplement Date Received by Manufacturer05/30/2019
Initial Report FDA Received Date02/20/2019
Supplement Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-