Model Number 8393.1814
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Rwmic considers this case open.The manufacture and user facility will be contacted to collect missing information.A follow up report will be submitted upon receipt of new or additional information.
 
Event or Problem Description
On january 21st 2019, the user facility reported the following to richard wolf medical instruments corporation (rwmic): while being used on a procedure, a part of the forceps fell in the patient.Will the device be returned? yes.Was the device being used during a procedure when the issue occurred? yes.Specifically, was the device being used on a patient when the issue occurred? yes.Was there any injury or illness to patient or other personnel due to issue? no.Did the issue cause a delay in the procedure being performed that put the patient at risk? yes.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.How was the patient anesthetized? unknown.
 
Additional Manufacturer Narrative
Follow up report # 1 is to provide fda with new and changed information.New / changed information : b3, d4, d10, f7, f11, f13, g4, g7, h2, h10 richard wolf medical instruments corporation (rwmic) considers this case open.Upon device evaluation or any additional information a follow up report will be submitted to fda.
 
Event or Problem Description
Follow up report to provide new and/or changed information.
 
Additional Manufacturer Narrative
Device labeling was reviewed for patient and device codes: patient code: not applicable, no patient problem was reported.Device code: ifu was reviewed the following cautions and checks were found: caution! excessive force can lead to breakage of or damage to the products.Due to the small dimensions required, the products have only limited strength.Use these instruments only to grasp and ablate small portions of tissue of the kinds defined in chapter 1.Immediately after use check the instruments for damage and for parts which may have broken off.Ensure that these parts do not remain in the patient.Caution! the products have only limited strength! exerting excessive force will cause damage, impair the function and therefore endanger the patient.Immediately before and after each use, check the products for damage, loose parts and completeness.Ensure that no missing instrument parts remain in the patient.Do not use products which are damaged, incomplete or have loose parts.Check instruments and accessories for damage, sharp edges and rough surfaces.Check jaw inserts for defective cutting edges and corroded parts and replace, if necessary.Check insulation of hf cable and sheaths and replace, if necessary.Caution! check for surface changes (e.G.Hairline cracks) in the area of the hinge pin the pin may loosen if the surface is damaged.Additional attempts were made to collect additional/missing information from the user facility, but no additional information could be provided.Rwmic considers this case closed.Should additional information become available a follow up report will be submitted.
 
Event or Problem Description
The user facility returned the device to richard wolf medical instruments corporation (rwmic) on june 26, 2019 and the device evaluation was completed on july 17, 2019.The device was visually and functionally evaluated using a magnifier.The instrument was found to have been repaired by a third party; as a result, the reported condition could not be verified.A probable root cause could not be determined as the instrument was repaired by a third party.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULINE
Common Device Name
ATRAUMATIC GRASPER FORCEPS MONOPOLAR
Manufacturer (Section D)
RICHARD WOLF GMBH
pforzheimer street 32
knittlingen, 75438
GM  75438
MDR Report Key8353592
Report Number1418479-2019-00003
Device Sequence Number11294445
Product Code GCJ
UDI-Device Identifier04055207062098
UDI-Public04055207062098
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K935270
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/21/2019
Report Date (Section F) 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8393.1814
Device Catalogue Number8393.1814
Device Lot Number112N11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/21/2019
Device Age8 YR
Event Location Hospital
Date Report to Manufacturer08/16/2019
Initial Date Received by Manufacturer 01/25/2019
Supplement Date Received by Manufacturer03/11/2019
07/17/2019
Initial Report FDA Received Date02/20/2019
Supplement Report FDA Received Date04/10/2019
08/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA08/16/2019
-
-