Model Number ACU0T0 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 02/01/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause.Additional information has been requested.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A health professional reported that during an intraocular lens (iol) implant surgery, the lens presented with a straight leading haptic.The lens sank into the vitreous after insertion into the eye.Additional information has been requested.
|
|
Manufacturer Narrative
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A non-qualified viscoelastic was indicated.The root cause may be related to a failure to follow the dfu.A non-qualified viscoelastic was used in the device.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.As stated in the company aspheric intraocular lens with company preloaded delivery system dfu, only qualified company viscoelastics must be used in conjunction with the device.All other viscoelastics are not qualified for use.Lens delivery performance may be negatively affected when using other non-qualified viscoelastics leading to lens delivery issues and /or damage.The use of non-qualified viscoelastics is considered a failure to follow the dfu for the company aspheric intraocular lens with company preloaded delivery system and is not recommended under any circumstance.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
New information received from the customer stating the surgery was completed after exchanging the lens.
|
|