The device was returned to olympus for evaluation.A visual inspection was performed on the returned device and found the bending section broken at approximately 75mm from the distal end.Prior to removal of the bending section cover, it was verified that there were no signs of the bending section skeleton protruding.The bending section cover was then removed to determine the reason for the bending section irregularity.Upon removal of the bending section cover, it was discovered that the bending section skeleton was partially broken with sharp edges adjacent to the insertion tube.The bending section skeleton tab is also bent causing it to lift at the same location where the bending section is broken.Additionally, an olympus "boroscope" was used to inspect the channel and noted numerous tear marks located within the bending section area, which also included a section of the biopsy channel that was found to be deformed.The scope¿s image was also inspected and found ig damage in the form of excessive broken fibers which cover 80 percent of the image and are within the field of view.A review of the scope's instrument history record showed the scope was purchased on (b)(6) 2018 and has not been returned for service/repair since the date of purchase.Based on similar reports related to the reported event and reported device, the likely cause of the could be attributed to the operator¿s technique.The instruction manual states the following; ¿do not twist or bend the bending section with your hands.Equipment damage may result¿.Additionally, the instructions for safe use manual indicate that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section; this would occur more so in a narrow environment.
|
Olympus was informed that during a diagnostic urology procedure, the scope¿s bending section broke.There was no bleeding observed.The intended procedure was completed with the same device.The patient was discharged home in good condition.Additionally, the user facility reported that the scope was inspected post procedure by the reprocessing technician with no visual deformation noted on the bending section of the scope.
|