Catalog Number 367341
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 02/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device type: jka, fpa.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that here was a needle sick injury with bd vacutainer® push button blood collection set, post use."clinician wasn't sure how it happened but knows it was after use, either during the disposable or immediately after the activation.Clinician sought treatment from employee health after the needle stick incident.".
 
Event or Problem Description
It was reported that here was a needle sick injury with bd vacutainer® push button blood collection set, post use."clinician wasn't sure how it happened but knows it was after use, either during the disposable or immediately after the activation.Clinician sought treatment from employee health after the needle stick incident.".
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
MDR Report Key8356054
Report Number1024879-2019-00261
Device Sequence Number13487673
Product Code JKA
UDI-Device Identifier50382903673419
UDI-Public50382903673419
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K030573
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2020
Device Catalogue Number367341
Device Lot Number8240820
Initial Date Received by Manufacturer 02/01/2019
Supplement Date Received by Manufacturer02/01/2019
Initial Report FDA Received Date02/20/2019
Supplement Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-