• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US BARREL BURR 4.0MM 5PK;  ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US BARREL BURR 4.0MM 5PK;  ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE Back to Search Results
Catalog Number 283469
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported by the affiliate via cst tool that during an unknown procedure the burr stopped rotating.The procedure was completed with no patient harm or time delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: the complaint device was received and inspected.The complaint can be confirmed.An attempt was made to rotate the inner shaft inside the outer shaft, however it was observed that the inner and outer shafts were jammed.When proper fluid management is not maintained while using the device, friction between the two shafts can cause heat to build up between the inner and outer shafts.Also, tissue debris can build up between the inner and outer shafts when proper fluid management is not maintained.When excessive tissue builds up between the two shafts, the shafts can become jammed.However, given the information provided we cannot discern a definitive root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance was identified.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Further it was reported that it was rotator cuff repair procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARREL BURR 4.0MM 5PK
Common Device Name
 ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8356221
Report Number1221934-2019-56452
Device Sequence Number16695095
Product Code HRX
UDI-Device Identifier10886705022045
UDI-Public10886705022045
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K041824
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number283469
Device Lot NumberM1811024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2019
Initial Date Received by Manufacturer 01/24/2019
Supplement Date Received by Manufacturer03/26/2019
Initial Report FDA Received Date02/20/2019
Supplement Report FDA Received Date04/17/2019
Patient Sequence Number1
-
-