| Model Number |
2429-0500 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, patient demographics not provided.Concomitant medical products: 250ml baxter bag, lot dr18f15096, intravia container.The customer¿s report of a leak was confirmed.Functional testing observed leakage from the top of the silicone segment.Closer inspection under magnification observed a leak due to a small hole on the silicone segment.No tool marks or crush marks were observed on the upper fitment near the small hole.The cause of the leak is a small hole damage in the silicone pump segment near the upper fitment.The root cause of the hole damage could not be definitively determined.
|
| |
|
Event or Problem Description
|
|
The set was received unexpectedly with a report that it leaked, potentially from a crack or flush.Although requested, no further information was received.
|
| |
|
Search Alerts/Recalls
|