Catalog Number 18063100
Health Effect - Clinical Codes Failure of Implant (1924); Device Embedded In Tissue or Plaque (3165)
Date of Event 11/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Customer reported that whilst using t2 recon nail, the registrar and scrub tested to see if they lined up with the jig.They took an x-ray and everything seemed ok, therefore proceeded to drill through the canal.When looking at the lateral view x-ray, they realized the screw missed the nail.They removed the 2 screws and adjusted the nail to try again, but the k-wire hit the nail and small piece of wire broke off.This was left in the patient and they decided not to continue this method.They successfully completed the surgery using antegrade technique by putting the screw through the lesser trochanter.There was a 45 minute delay.Delay of 45 minutes, unintended metal (k-wire) left in patient.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.This is the first reported case with this lot number.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
Customer reported that whilst using t2 recon nail, the registrar and scrub tested to see if they lined up with the jig.They took an x-ray and everything seemed ok, therefore proceeded to drill through the canal.When looking at the lateral view x-ray, they realized the screw missed the nail.They removed the 2 screws and adjusted the nail to try again, but the k-wire hit the nail and small piece of wire broke off.This was left in the patient and they decided not to continue this method.They successfully completed the surgery using antegrade technique by putting the screw through the lesser trochanter.There was a 45 minute delay.Delay of 45 minutes, unintended metal (k-wire) left in patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET DEVICE T2 RECON
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key8359171
Report Number0009610622-2019-00079
Device Sequence Number19168897
Product Code HSB
UDI-Device Identifier04546540654939
UDI-Public04546540654939
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
K102992
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number18063100
Device Lot NumberKME903313
Was Device Available for Evaluation? No
Date Returned to Manufacturer02/15/2019
Initial Date Received by Manufacturer 02/06/2019
Supplement Date Received by Manufacturer04/09/2019
Initial Report FDA Received Date02/21/2019
Supplement Report FDA Received Date05/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1