The reported event was confirmed as manufacturing related based on the evaluation.Ten stock samples were received for evaluation.The samples were received open and with original packaging.Visual observation noted a fuzzy eye at the proximal end of the eyelet.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿caution: this product contains natural rubber latex which may cause allergic reactions.With funnel end 16¿ (40.6cm) length, two eyes, x-ray opaque rubber warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|