| Model Number |
MMT-751PNAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had keypad unresponsive.The customer blood glucose level was 140 mg/dl.The customer was not assisted with troubleshooting.Customer stated that the insulin pump lost settings when changing out batteries and insulin pump had problems reading new batteries.Customer stated that the buttons were not working properly.Customer stated that the rewind seems slower and scratches on screen and customer experiencing no delivery alarms without explanation.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with no button response at bolus, esc, act, up arrow and down arrow button due to unlocked j2 connector on lcd board noted.
|
| |
|