Catalog Number 362800
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that a bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg partially filled.
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for underfill with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for underfill with the incident lot was observed.Root cause description: based on the investigation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that a bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg partially filled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 15.8 MG
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8363209
Report Number2243072-2019-00315
Device Sequence Number1588044
Product Code JKA
UDI-Device Identifier50382903628006
UDI-Public50382903628006
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K972075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2019
Device Catalogue Number362800
Device Lot Number8215578
Initial Date Received by Manufacturer 02/04/2019
Supplement Date Received by Manufacturer02/04/2019
Initial Report FDA Received Date02/22/2019
Supplement Report FDA Received Date03/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-