|
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Unopened samples of product code y682h, lot lck083 were returned for analysis.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the sample received, no pull off suture needle were found, and the tested sample met the finished goods requirements.
|