| Catalog Number |
5959124 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Defective Device (2588); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Abscess (1690); Adhesion(s) (1695); Bacterial Infection (1735); Fistula (1862); Strangulation (2084); Hernia (2240); Injury (2348); Obstruction/Occlusion (2422)
|
| Date of Event |
08/08/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time no conclusions can be made to what extent the bard/davol sepramesh ip may have caused or contributed to the reported event.The cause of the patient postoperative complications cannot be determined at this time.The patient's attorney alleges subsequent surgical intervention; however, no details have been provided.No lot number has been provided therefore a review of the manufacturing records is not possible at this time.Should additional information be provided a supplemental emdr will be submitted.Not returned.
|
| |
|
Event or Problem Description
|
|
The following was alleged by the patient's attorney: (b)(6) 2011: the patient underwent surgery for implant of an unspecified bard/davol sepramesh ip.Ni/ni/ni: the patient had subsequent surgical intervention due to the hernia mesh device(s).The patient is making a claim for an adverse patient outcome against the bard/davol sepramesh ip.The attorney alleges patient experienced emotional distress and the device was defective.
|
| |
|
Event or Problem Description
|
|
The following was alleged by the patient's attorney: (b)(6) 2011: the patient underwent surgery for implant of an unspecified bard/davol sepramesh ip.Ni/ni/ni: the patient had subsequent surgical intervention due to the hernia mesh device(s).The patient is making a claim for an adverse patient outcome against the bard/davol sepramesh ip.The attorney alleges patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2011 - patient was diagnosed with incarcerated incisional hernia thereby underwent open repair with the implant of sepramesh ip.Per operative notes, ¿multiple other small incisional hernias were found.The bowel was free of adhesions and a sepramesh ip was placed and sutured." (b)(6) 2015 - patient visited hospital and was diagnosed with small bowel obstruction.(b)(6) 2015 - patient was diagnosed with strangulated incisional hernia, umbilical hernia, small bowel obstruction thereby underwent open repair with implant of absorbable synthetic mesh.Per operative notes, ¿adhesions were lysed in all the four quadrants, the edge of the mesh was found once had reduced all the hernias.A chronic draining tract with suture abscess was found.The recurrent incisional hernia was repaired, and an absorbable synthetic mesh was placed underlay and sutured.Then, the umbilical hernia defect was repaired, and an absorbable synthetic mesh was placed underlay and sutured." (b)(6) 2015 - patient was diagnosed with abdominal wall abscess.(b)(6) 2016 - patient was diagnosed with incarcerated incisional hernia, small bowel obstruction.(b)(6) 2016 - patient was diagnosed with infected mesh exposed completely through the skin, recurrent incisional hernia thereby underwent open repair with removal of infected mesh and implant of absorbable synthetic mesh.Per operative notes, ¿there were multiple drainage site with obvious exposed synthetic mesh was identified, numerous fistulas sites were excised along with all the old synthetic meshes.Some necrotic, atretic muscle and fascia was also excised in addition to the hernia sac.An absorbable synthetic mesh was placed as an underlay and sutured.Next, the fascial edges were reapproximated with sutures and an absorbable synthetic mesh was placed as an onlay and sutured." attorney alleged that the patient had abscess, adhesions, bowel obstruction, fistulae, infection, mesh migration, ring break, seroma, hernia recurrence, pain & suffering.It was also alleged that the mesh migrated and protruded through skin, mesh infection led to subsequent surgeries and further damage to kidneys.
|
| |
|
Additional Manufacturer Narrative
|
|
At this time no conclusions can be made to what extent the bard/davol sepramesh ip may have caused or contributed to the reported event.The cause of the patient postoperative complications cannot be determined at this time.The patient's attorney alleges subsequent surgical intervention; however, no details have been provided.No lot number has been provided therefore a review of the manufacturing records is not possible at this time.Addendum: this supplemental emdr is submitted to document additional information provided.Based on the additional information received, there is no change to initial determination, no conclusions can be made.Per medical records review, about 2 years 4 months post implant of sepramesh ip, patient was diagnosed with strangulation, fistula, bacterial infection, abscess, obstruction, hernia recurrence and adhesions thereby underwent repair with removal of mesh.The instructions-for-use supplied with the device lists hernia recurrence, fistula and adhesions as possible complications.In regard to infection, the warnings section in ifu states, "if an infection develops, treat the infection aggressively.Consideration should be given regarding the need to remove the prosthesis.An unresolved infection may require removal of the prosthesis." a review of the manufacturing records was performed and found that the lot was manufactured to specification.Note, the manufacturing date (15-aug-2010) is considered to be a best estimate.Updated fields: a2, a4, b2, b4, b2, b3, b4, b5, b6, b7, d4 (catalog no., corporate lot no, udi, expiry), d6.B (explant), g1, g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|